Pacific Instruments, Inc. Pacific Instruments, Inc.
4080 Pike Lane
Concord, CA 94520
www.pacificinstruments.com

(925) 827-9010
salesrequest@pacificinstruments.com
Products > Turnkey Acquisition Systems > Turnkey Laboratory Acquisition Systems > Model 6825  


Model 6825, Turnkey ICP/IEPE Data Acquisition System


Features

  • 16 Channels ICP/IEPE/Voltage Input
  • 200kS/s 16-Bit Digitizer
  • High Level Buffered/Calibrated Analog Outputs
  • AC/DC Coupling
  • BNC Inputs
  • IRIG Time A, B or G
  • Digital I/O
  • PI660 Turnkey Software Included

Description

Series 6800 is a ready-to-run, high performance laboratory based Measurement System. Enclosures are 3U rack mountable or can be used right on the desktop. 6800 units come equipped with a conditioner, amplifier, filter and digitizer. Each channel has a front mounted BNC connector for voltage or ICP/IEPE inputs. Individual channels have a differential +/- 10V input and can digitize and record data to 16-bit resolution with upper and lower programmable alarm capability that is checked each time the output is digitized. Solutions are available for providing digital outputs based on these alarm conditions.

Systems interface to a familiar Windows based operator workstation over USB 2.0 where the included software, PI660, is used for setup, display & acquisition of the measurement data. PI660 allows for acquisition and control of one or many 6800 systems and provides real time data display. Data can be displayed and recorded in any number of ways, creating a versatile system that can fit any application. Series 6800 is fully compatible with Pacific Instruments’ Series 6000 Transducer Data Acquisition Systems. Custom configurations are easily accommodated with this system; please contact our sales engineers for additional information.







Overview  · Input  · Excitation/Transducer Power  · Amplifier  · Filter  · Digitizer  · Calibration  · Operation  · Data Format  · Data Interface  · Connections  · Mechanical  · Control & Data Processor (CDP) and Operators Workstation (PCCOWU) (Optional)  · Ordering Information 

Overview

Measurement Type

4-20mA
ICP/IEPE

Insrumentation Amp

Voltage

Analog Bandwidth

100 kHz

Excitation/Transducer Power

+/-12V
+/-15V

24V

ICP

Filter

4-Pole
6-Pole

8-Pole

Sample Rate

Up to 200 kS/s


Input

Configuration

8 channels, differential, 2-wire with shield.

Input Type

Programmable AC or DC input. Input attenuator and current input are available.

Range

±2 mV to ±10 Volts.

Impedance (AC)

100k Ohms, shunted by 1,000 pF.

Impedance (DC)

50 Megohms, shunted by 500 pF.

Protection

±50 Volts differential, ±30 Volts common mode.


Excitation/Transducer Power

Current

2 to 20 mA. 6 mA is supplied unless otherwise specified.

Minimum Compliance

24 V

Verification

Short and open detection.

Voltage

±12 or ±15 Volts jumper selectable per channel, ±24 also available.


Amplifier

Gain

Programmable 1 to 5000, in 1, 2, 3, 5 steps, with ±0.05% accuracy

Gain Stability

±0.01%, ±0.005%/ºC.

Linearity (Amplifier)

±0.01% for Gains < 1,000, ±0.02% for gains > 1,000.

Common Mode

60 dB plus gain in dB to 110 dB, DC to 60 Hz.

CM Voltage

±10 V

Zero

Automatic to ±1 mV.

Zero Stability X1

±1 mV, ±0.2 mV/ºC

Zero Stability X1000

±5 mV, ±1 mV/ºC

Noise X1

0.2 mV RMS for 20 kHz bandwidth.

Noise X1000

2.8 mV RMS for 20 kHz bandwidth.

Bandwidth

DC to 100 kHz (-3 dB). 1 Hz to 100 kHz (-3 dB) in AC coupled mode.

Slew Rate

3.2 V/µS

Analog Output (Amplifier)

±10 Volts full scale, 20 mA. Programmable for wideband or filtered response.
Buffered Analog Output.


Filter

Type

Single frequency four-pole (standard) or 4 frequency four-pole or 2 frequency eight-pole Bessel.

Frequency

From 250 Hz to 50 kHz
Standard Four Pole: 10 kHz
4 Frequency Four Pole: 10 Hz, 1 kHz, 10 kHz, 20 kHz
2 Frequency Eight Pole: 2 kHz and 20 kHz.

Noise

1 mV peak, RTO.

Other

Other filter characteristics and cut offs are available.


Digitizer

Sample

±50 ns channel-to-channel time correlation.

Resolution

16-bits, two's complement output per channel.

Rate

Programmable up to 200 kS/s digitizer per channel.

Digitizer Linearity

±1-1/2 LSB (±0.004%)

Continuity

Monotonic to 15 bits.


Calibration

Voltage Substitution

Alternate input for external calibration source. Programmable 1, 0.1 and 0.01, attenuation with ±0.02% accuracy. Attenuator output may be connected to output bus for accuracy check.

Zero Calibration

Amplifier input disconnected and shorted for zero calibration.


Operation

Alarms

Warning and alarm buses may be independent or shared between enclosures and may initiate an output from a digital I/O type module.
Two alarms each with upper and lower limits that are programmable from negative to positive full scale. Limits checked on each ADC sample.

Protocol

Control and data interface is USB 2.0.

Software

Window's driver (Both 32 and 64-bit) provides high-level operating command set. Fully compatible with all implementations of PI660 operating software.

Control Inputs

TTL inputs for Start, Stop and Trigger assert flags in the header of output data that initiate software control operations.


Data Format

Scan Table

Maximum format length is 65,536 samples.

Sample Rates

Multiple sample rates consisting of the highest sample rate divided by binary numbers. Highest sample rate is programmable with 1 uS resolution.

Data Word

16 or 32-bits, 2's complement binary.


Data Interface

Output Rate

Processor dependent, typically over 5 million 16-bit samples/second.

Latency

Processor and scan table dependent, typically less than 5 milliseconds.

Clock Stability

100 ppm over temperature range.


Connections

Calibration

15-Pin Type D mounted on rear panel. Mating connector supplied.

Control

15-Pin Type D mounted on rear panel. Mating connector supplied.

Synchronization

Category 5, enclosure to enclosure. 2-meter cable supplied.

USB

2-meter cable supplied.


Mechanical

Power (Mechanical)

115 or 230 VAC, 110%, 47 to 63 Hz, 400 Watts.

Operating Temperature (Mechanical)

0 to +50 ºC

Humidity

95% without condensation.

Shock/Vibration

Normal shipping and handling of laboratory instruments.

Width

View
19 in

Height

View
5 in

Depth

View
18 in

Weight

Approximately 25 pounds with all channel modules.

Mounting

Occupies one slot in Series 6000 enclosures.

Connectors

Input connector is 50-pin Type D. Output is 15-pin Type D High Density. Mating connectors are supplied.

Temperature

0 to +50 ºC


Control & Data Processor (CDP) and Operators Workstation (PCCOWU) (Optional)

Input

USB 2.0 ports for up to eight 6000U enclosures. Additional 6000U enclosures accommodated by adding USB expansion cards.


Ordering Information 

Ordering Information

6825-16 - 16-Ch ICP/IEPE/Volatge DAS w/ Analog output (BNC)

Note

-

Pacific Instruments, Inc. · 4080 Pike Lane · Concord, CA 94520 · www.pacificinstruments.com