Model 6860, 3U Thermocouple Scanner
|
|
|
|
Features
- 32, 64 or 96 Channel Configurations in a 3U Rack Mount Enclosure
- Utilizes External Uniform Temperature References (UTRs)
- UTR Model 6084-32, Screw Terminal Reference Allows for Integration of Any TC type and/or
- UTR Model 6084-32-RM is a 2U, Rack Mountable UTR with Front Mounted 2-pin Mini TC Jacks (T, K, E or J)
- IRIG Time - A, B or G
- PI660 Turnkey Software Included
Description
Series 6800 ready-to-run, high performance Thermocouple Measurement Systems are available in many configurations. Each 6800 is equipped with an amplifier, filter and digitizer plus a Uniform Temperature Reference (UTR). Systems interface to a familiar Windows based operator workstation over USB 2.0 where the included software, PI660, is used for setup, display and acquisition of thermocouple data. Each channel has a differential +/- 10V input and can digitize and record data to 16-bit resolution. Digitization rates up to 2,000 samples per second per channel are provided. Additionally, individual channels have an upper and lower programmable alarm capability that is checked each time the output is digitized. Solutions are available for providing digital outputs based on these alarm conditions. PI660 allows for acquisition and control of one or many 6800 systems and provides real time display of thermocouple data. Data can be displayed and recorded in any number of ways, creating a versatile system that can fit any application. Series 6800 is fully compatible with Pacific Instruments’ Series 6000 Transducer Data Acquisition Systems.
PI660 Turnkey Software
PI660 software is the choice for test management, data acquisition, and data display. PI660 controls the Pacific Instruments Series 6800 Data Acquisition System. It provides all the tools for test definition, test setup and tracking, transducer calibration, data acquisition and display, data replay, data export, and test quality validation. PI660 is turnkey, intuitive and easy to use. The ribbon style control panel provides logical access to the various steps of programming. The software even supports a password protected user account mode where functionality can be turned on or off based on the needs of the facility. Learn more about PI660 Turnkey Software.
|
|
|
|
|
|
|
Configuration
|
|
32 channels differential, 2-wire with shield.
|
|
Range
|
|
±2 millivolts to ±10 Volts full scale.
|
|
Impedance
|
|
50 Megohms, shunted by 1,000 pF.
|
|
Protection
|
|
±25 Volts differential and common mode.
|
|
Type (Thermocouple Inputs)
|
|
B, C, E, J, K, N, R, S, and T.
|
|
Configuration (Thermocouple Inputs)
|
|
Differential, 2 wire with shield.
|
|
|
|
|
Gain
|
|
Programmable 1-5000, in 1, 2, 3, 5 steps, with ±0.05% accuracy.
|
|
Gain Stability
|
|
±0.01%, ±0.005%/ºC.
|
|
Linearity (Amplifier)
|
|
±0.01% for gains < 1,000, ±0.02% for gains 1,000 and higher.
|
|
Common Mode
|
|
75 dB plus gain in dB to 110 dB, DC to 60 Hz.
|
|
CM Voltage
|
|
±10 V
|
|
Source Current
|
|
±25 nA, ±0.1 nA/ºC.
|
|
Zero
|
|
Automatic to ±1 µV RTI, ±0.5 mV RTO.
|
|
Zero Stability
|
|
±5 µV RTI, ±1 mV RTO, ±1 µV/ºC RTI, ±0.2 mV/ºC RTO. Short term: ±2 µV RTI, ±0.4 mV RTO for 8 hours.
|
|
Noise (1 kHz)
|
|
2 µV RTI plus 0.55 mV RTO, RMS
|
|
Crosstalk
|
|
Below -86 dB for full scale signal on adjacent channels.
|
|
Bandwidth
|
|
Determined by input filter, 1 kHz (-3 dB) maximum.
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Input
|
|
USB 2.0 ports for up to eight 6000U enclosures. Additional 6000U enclosures accommodated by adding USB expansion cards.
|
|
|
|
|
Ordering Information
|
|
6860-128 - 128-Ch TC Scanner w/ external UTR 6860-32 - 32-Ch TC Scanner w/ external UTR 6860-64 - 64-Ch TC Scanner w/ external UTR 6860-96 - 96-Ch TC Scanner w/ external UTR
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|