Model 6013, 8-Channel Instrumentation Amplifier-Filter-Digitizer


Features

  • Voltage, thermocouple & DC-LVDT inputs
  • Optional thermocouple reference junction box
  • Gains 1 to 5,000 with 0.05% accuracy
  • Automatic zero & gain calibration
  • Four-pole, low-pass filter
  • Up to 10 kS/s per channel with 16-bit resolution
  • Two alarms with programmable upper & lower limits

Description

The 6013 input module has eight channels, each with programmable gain instrumentation amplifier, low pass filter and sample and hold. The high level outputs are multiplexed and digitized to 16 bits then output to the 6000 data bus. A ninth reference temperature channel conditions the output of the temperature sensor in the Model 6084 thermocouple reference junction box The 6013 provides regulated DC power for transducers with integral electronics. Each channel has a wideband analog output.

The 6013 may be configured for use with voltage, thermocouple and current transducers. It provides ±12/15 Volt power for transducers with integral electronics. Versions are available for current loop, high voltage and 28 Volt powered transducers.

Voltage substitution is provided for channel gain calibration utilizing an external voltage standard. A calibration attenuator enables the voltage standard to be used on its highest accuracy ranges and provides a post-attenuator output for calibration and verification. Using Pacific's PI660 software zero and gain calibration and correction are automatic.

The four-pole, low-pass filter uses an easily changed plug-in module to set bandwidth. Either the wideband or filtered output may be digitized and sent to the 6000 data bus. Two programmable alarms each with upper and lower limits are checked each time the outputs are digitized.







Overview  · Input  · Excitation/Transducer Power  · Amplifier  · Filter  · Digitizer  · Calibration  · Mechanical  · Ordering Information 

Overview

Measurement Type

4-20mA
Position

Thermocouple

Voltage

Analog Bandwidth

1 kHz

Excitation/Transducer Power

±12V
±15V

24V

Filter

4-Pole

Sample Rate (Digitizer)

Up to 10 kS/s

Analog Output

Monitor Output


Input

Configuration (Input)

8 channels, differential, 2 wire with shield.

Range

±2 mV to ±10 Volts (±200 mV to ±100 V with optional attenuator).

Attenuator

100:1, ±0.2% (6013-HV).

Impedance

50 Megohms, shunted by 1,000 pF (1 Megohm with attenuator)

Protection

±50 Volts differential, ±30 Volts common mode. ±250 Volts differential with attenuator.

Type (Thermocouple Inputs)

B, C, E, J, K, N, R, S, and T.

Configuration (Thermocouple Inputs)

Differential, 2 wire with shield.

Type (Current Inputs)

4-20 mA input.

Configuration (Current Inputs)

Differential, 2 wire with shield.

Termination (Current Inputs)

200 Ohms


Excitation/Transducer Power

Voltage

Regulated ±12 or ±15 Volts jumper selectable per channel (6013). 28 Volts (6013-24V).

Current

50 mA per channel, limited to 200 mA maximum per 8-channel module.


Amplifier

Gain

Programmable 1-5000, in 1, 2, 3, 5 steps, with ±0.05% accuracy.

Gain Stability

±0.01%, ±0.005%/ºC.

Bandwidth

1 kHz
-3
dB

Linearity (Amplifier)

±0.01% for gains 0.1 to 1,000, ±0.02% for gains above 1,000.

Common Mode

60 dB plus gain in dB to 110 dB, DC to 60 Hz.

CM Voltage

±10 V

Zero (Amplifier)

Automatic to ±1 µV RTI, ±0.5 mV RTO.

Zero Stability

±5 µV RTI, ±1 mV RTO. ±1 µV/ºC RTI,
±0.2 mV/ºC RTO. Short term: ±2 µV RTI,
±0.4 mV RTO for 8 hours.

Source Current

±10 nA, ±1 nA/ºC.

Noise (10 Hz)

0.1 µV rms, RTI, 0.5 mV rms, RTO

Noise (1 kHz)

1.0 µV rms, RTI, 0.5 mV rms, RTO

Recovery

800 µS to ±0.1% for 10X overload to ±10 V.

Monitor Output

±3.0 Volts full scale, unfiltered.


Filter

Type (Filter)

Four-pole, low-pass Butterworth.

Frequency

Plug-in, 4 Hz to 1 kHz, 10 Hz supplied. Alternate fiter characteristics and frequencies are available.

Other

Other filter characteristics and cut offs are available.


Digitizer

Sample

Simultaneous sample and hold with ±50 nS channel-to-channel. Droop is less than ±0.005%.

Resolution

16-bits, two's complement output.

Sample Rate

Up to 10 kS/s per channel
10
kS/s

Digitizer Linearity

±2 LSB (±0.006%)

Continuity

Monotonic to 15 bits.

Alarms

Two alarms each with upper and lower limits that are programmable from negative to positive full scale. Limits checked on each ADC sample.



Calibration

Voltage Substitution

Alternate input for external voltage standard. Programmable attenuation steps of 1, 0.1, and 0.01 with ±0.02% accuracy. Output of the attenuator is provided on a rear panel connector for calibration.

Zero Calibration

Amplifier input disconnected and shorted.


Mechanical

Mounting

Occupies one slot in Series 6000 enclosures.

Connectors

Input connector is 50-pin Type D. Connectors are mounted on the front and mates are supplied.

Operating Temperature

0 to +50 ºC


Ordering Information 

Ordering Information

6013 - 8-Channel Voltage/Thermocouple, ±12V or ±15V Power.


6013-24V - 8-Channel Voltage/Thermocouple 28V Power.


6013-HV - 8-Channel Voltage, 100:1 Attenuator.


6013-I - 8-Channel Current Loop. 28 Volts, 0 to 20mA.


6081 - Screw Terminal Adapter.


6084 - 8-Channel Thermocouple Reference Junction.