Model 7200, Ruggedized 16-Channel Strain/Bridge/Voltage Data Acquisition
|
|
|
|
Features- 16-Channels strain gage, bridge or voltage transducers
- Ethernet control and data interface
- Real time temperature correction of Gain, Zero & Excitation from -20°C to +50°C
- 0.05% or better accuracy over full gain & temperature range
- 10 kHz analog bandwidth, gain 1 to 1,000
- 24-Bit A/D programmable up to 24k S/s per channel
- Wide range programmable filter, 80 dB/octave
- IRIG A, B & G time decoding and recording
- 1 μS synchronized sampling of multiple enclosures
- Ruggedized for tough environments
- 20 to 30 Volt DC power
DescriptionPacific’s Ruggedized Series 7200 brings a brand new look to strain gage and transducer data acquisition. A network appliance, each module conditions, amplifies and digitizes the outputs of up to sixteen strain gage, bridge or voltage transducers. The rugged enclosure is designed for installation in wind tunnels, engine test stands and other facilities where it is desired to locate the DAS close to the test article, ultimately reducing installation and cable costs as well as improving signal quality and reliability. Fully self-contained, the 7200 provides transducer excitation, bridge completion and balance. The transducer output is amplified, filtered to remove signal components that could produce alias errors and digitized to 24-bit resolution. The digitized transducer output is digitally processed providing multiple sample rates, an 80 dB/octave FIR filter and 32-bit floating point output format in millivolts referred to input, millivolts referred to output or user defined engineering units. Four models are available that differ in their excitation and signal conditioning capabilities. Model 7214A and 7216A for strain gage and bridge transducers have programmable voltage excitation, bridge completion for 120 and 350 Ohm strain gages, shunt calibration and automatic balance. Model 7220A is similar but provides regulated power for transducers with internal electronics. Model 7218A is for voltage inputs. The amplified transducer signals are digitized, processed and output at 10 to 24,000 samples per second, as selected for individual channels. A digital signal processor (DSP) provides low-pass filtering selectable from 10 Hz to 3,000 Hz with 80 db/octave roll-off. Each module has a 1 Gigabit Ethernet interface for control and data output, which simplifies system wiring and provides remote and distributed operation. IRIG time may be employed to synchronize sampling on multiple modules within 10 microseconds, or a TTL timing may be used to achieve better than 1 microsecond time alignment. Multiple modules with up to 4,096 channels may be connected on a single Ethernet LAN. The module enclosure features interchangeable end plates that provide fully sealed, ambient air or forced air cooling. Forced air cooling provides the capability to operate over a wide temperature range according to application requirements. Occupying less than 500 cubic inches and weighing less than 14 pounds, the 7200 modules can be installed in locations previously not suitable for data acquisition systems. The Ethernet interface and IRIG timing make it easy to distribute Bricks throughout the test facility, minimizing sensor cabling and installation costs. The 7200 does not sacrifice performance for size or portability; it features higher accuracy than most systems designed for a laboratory environment. For the 7216, accuracy is better than 0.025% for gains up to 200 and 0.05% for gains above 200 over the temperature range -20°C to +50°C, which can be extended to higher temperatures by forced air cooling. Series 7200 is supported by Pacific’s PI770 software that provides a ready-to-run application and an SDK for user program development. Calibration and performance verification is accomplished by PANEL72 Maintenance and Calibration software. The ACS2000 Automatic Calibration System automates performance verification and periodic instrument calibration.
|
|
|
|
|
Configuration
|
|
7220A: Four-wire input, signal input (2), transducer power (2) and shield. 7214A & 7216A: Six-wire input, signal input (2), excitation (2), excitation sense (2) and shield. Programmable bridge completion for half bridges and 120 Ohm and 350 Ohm quarter bridges. 7220A: Four-wire input, signal input (2), transducer power (2) and shield.
|
|
Input Stability
|
|
±0.02% for 8 hours, ±0.005%/ºC.
|
|
Input Impedance
|
|
50 Megohms shunted by 500 pF.
|
|
Input Protection
|
|
±50 Volts differential, ±50 Volts common mode without damage.
|
|
Bridge Balance
|
|
7216A: Automatic by program control. Balance accuracy ±0.025% of range, ±1 mV RTO. Range set by resistor up to 25 mV/V, 2.5 mV/V (350 Ohms) installed.
|
|
|
|
|
|
|
Input Range
|
|
±10 mV to ±10 Volts.
|
|
Gain
|
|
Programmable from 1 to 1,000 with 0.02% resolution. Thirteen calibrated gain steps are provided: 1, 2, 3, 5, 10, 20, 30, 50, 100, 200, 300, 500, 1000 with ±0.02% accuracy.
|
|
Gain Stability
|
|
±0.01% for 30 days, ±0.001%/ºC compensated. ±0.05% for 30 days, ±0.005%/ºC without compensation.
|
|
Gain Linearity
|
|
±0.02 %
|
|
Common Mode
|
|
60 dB plus gain in dB to 110 dB, DC to 60 Hz.
|
|
Operating CM Voltage
|
|
±10 V
|
|
Zero (Amplifier)
|
|
Automatic zero to ±2 µV RTI or ±1.0 mV RTO whichever is greater.
|
|
Zero Stability
|
|
±2 µV RTI plus ±1 mV RTO over the full temperature range.
|
|
Source Current
|
|
±40 nA, ±0.4 nA/°C
|
|
Noise (10 Hz)
|
|
2.0 µV RTI plus 0.3 mV RTO, RMS.
|
|
Bandwidth
|
|
10 kHz (-3dB)
|
|
Overload
|
|
Recovery time is 120 µS to within ±0.1% for a 10 times overload to ±10 Volts.
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Ordering Information
|
|
7218A - 16-Ch Voltage Data Acquisition 7214A - 16-Ch Strain/Bridge Data Acquisition 7218A - 16-Ch Voltage Data Acquisition 7220A - 16-Ch Transducer Data Acquisition
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|