Model 584X, Transient Recording I/O Modules
|
|
|
|
Features
- Record fast events, up to 10 MS/s
- One pre-trigger and two post-trigger sample rates
- Rugged 3 or 16 channel, battery operated enclosures
- 16 Channel A/C powered laboratory enclosures
- Ethernet interface with Windows software
- Dual inputs, bridge and ICP
- Programmable for 1/4, 1/2 and full bridges
- Voltage & shunt calibration
- Automatic, hands-off operation
- Analog output for external recording
Description
Individual channel modules that plug into the 5800 enclosures condition, amplify, filter, digitize and store the transducer signals. Simultaneous sampling provides excellent time correlation between channels and external events. The digitizer is available with 14, 16 or 24-bit resolution and sample rates up to 10 Million samples per second.
Data acquired by the 5800 is digitized and recorded in non-volatile, solid-state memory. It records calibration, pre-trigger and post-trigger data. In a typical test scenario recording of pre-trigger data is initiated by a TTL input, program instruction, or automatic sequencer. A TTL input or on-board discriminator triggers post trigger data recording.
The quantities and sample rates of pre and post-trigger data storage are programmable, enabling the 5800 to be configured according to the characteristics of the data being acquired. Sample rate can be changed during acquisition without interrupting data flow to capture high-speed events, yet have sufficient memory space to record long-term effects. The 5800 can be configured for multiple records, each using the same pre and post-trigger sampling profile. Data is debriefed and transferred to permanent storage by the Ethernet interface.
Each channel has two inputs: one for bridges, voltage, RTDs, potentiometers and thermocouples; the other AC coupled with current excitation for transducers with built-in electronics. Signal conditioning includes programmable excitation and completion for 1/4, 1/2 and full bridge, AC-coupling, automatic balance, voltage and four-step shunt calibration, programmable gain and a four frequency, eight-pole filter.
|
|
|
|
|
|
|
|
|
|
|
|
|
Input Range
|
|
±10 mV to ±10 Volts full scale.
|
|
Gain
|
|
Programmable steps 1, 2, 3, 5, 10, 20, 30, 50, 100, 200, 300, 500 and 1,000.
|
|
Gain Stability
|
|
±0.02% for 8 hours, ±0.005%/ºC.
|
|
Gain Linearity
|
|
±0.02 %
|
|
Input Impedance
|
|
25 Megohms shunted by 500 pF.
|
|
Input Protection
|
|
±50 Volts differential or common mode.
|
|
Common Mode
|
|
60 dB plus gain in dB to 110 dB minimum for 350 Ohm source unbalance, DC to 60 Hz.
|
|
CM Voltage
|
|
±10 V
|
|
Zero Stability
|
|
±5 µV RTI, ±1 mV RTO at constant temperature. ±1 µV/ºC RTI, ±0.2 mV/ºC RTO.
|
|
Noise
|
|
10 µV RTI plus 1 mV RTO RMS.
|
|
Bandwidth
|
|
5841/43 - 100 kHz. Slew rate is 5 V/µS. 5842/44/45/46 - 500 kHz for gains below 100, 300 kHz for higher gains. Slew rate is 15 V/µS.
|
|
|
|
|
Filter Type
|
|
5841/42/44/45/46 - Eight-pole, Bessel, low-pass (48 dB/octave terminal slope). 5841/43 - Standard is 5 kHz, 10 kHz, 20 kHz, 50 kHz (-3 dB) and wideband. 5842/44/45/46 - Standard is 10 kHz, 20 kHz, 50 kHz, 100 kHz (-3 dB) and wideband. Other filter frequencies are available on request.
|
|
|
|
|
|
|
Sampling
|
|
One pre-trigger and two post-trigger sample rates per record.
|
|
Rates
|
|
5841: 1M, 500K, 200K, 100K, 50K, 20K, 10K, 5K, 2K, 1K, 500, 200, 100, 50, 20 and 10 S/s. 5842: 2M, 1M, 500K, 200K, 100K, 50K, 20K, 10K, 5K, 2K, 1K, 500, 200, 100, 50, and 20 S/s. 5843: 200K, 100K, 50K, 20K, 10K, 5K, 2K and 1K S/s. 5844/45/46: 10M, 5M, 2M, 1M, 500K, 200K, 100K, 50K, 20K, 10K, 5K, 2K, 1K, 500, 200 and 100 S/s.
|
|
Resolution
|
|
5843 - 24-bits 5841 & 5846 - 16-bits 5842, 5844 & 5845 - 14-bits
|
|
Correlation
|
|
1 nS sample to sample, 50 nS channel to channel.
|
|
Linearity
|
|
5841 - 3.5 LSB 5842, 5844, 5845 & 5846 - 2.5 LSB
|
|
|
|
|
|
|
|
|
Ordering Information
|
|
5841 - Transducer Recorder Module, 1 MS/s, 16-bit, 4M storage. 5842 - Transducer Recorder Module, 2 MS/s, 14-bit, 1M storage. 5843 - Transducer Recorder Module, 200 KS/s, 24-bit, 3.5M storage. 5844 - Transducer Recorder Module, 10 MS/s, 14-bit, 8M storage. 5845 - Voltage Recorder Module, 10 MS/s, 14-bit, 8M storage. 5846 - Transducer Recorder Module, 10 MS/s, 16-bit, 8M storage. 5847 - Voltage Recorder Module, 10 MS/s, 16-bit, 8M storage.
|
|
|
|
|
|
|
|
|
|
|
|
|
|