|
|
|
Model 6160, 2-Channel Transducer Amplifier-Filter-Analog Output, 300CMV
|
|
|
|
Features
- Isolated excitation & input with 300 Volts common mode
- Plug-in channel configuration & calibration card
- Voltage & current excitation with remote sensing
- Automatic zero & balance
- Voltage substitution, DAC or 2/4 step shunt calibration
- Gains 1 to 5,000 with 50 kHz or 100 kHz bandwidth
- Four six-pole low-pass filters, optional programmable filter
- Dual buffered 10 Volt analog outputs
Description
The 6160 is a two-channel signal conditioning amplifier module with 50 kHz or 100 kHz bandwidth and analog outputs. The bridge input is ten-wire shielded with programmable constant voltage or constant current excitation and programmable completion for quarter, half and full bridge transducers. Automatic bridge balancing accommodates large imbalances without limiting dynamic range.
The input and excitation are isolated from the outputs, power and control interface. This gives the user complete freedom to ground the input without creating ground loops that introduce noise and offset errors.
The differential instrumentation amplifier has programmable gains from 1 to 5,000 and automatic zero. The standard filter is a six-pole Bessel with four programmable bandwidths. An optional four-pole Bessel filter has continuously programmable bandwidth.
A “features card” provides shunt calibration using dedicated inputs. Two-step, resistive shunt calibration is standard. Four-step shunt calibration and simulated shunt using a DAC with 16-bit resolution are also available. Voltage substitution using an external source is provided for traceable gain calibration.
|
|
|
|
|
|
|
Configuration
|
|
Input configuration based on installed Features Cards. Features cards available for Bridge, IEPE and RTD. Other features cards available upon request.
|
|
|
|
|
|
|
|
|
|
|
Input Range
|
|
±2 mV to ±10 Volts full scale, DC or AC coupled.
|
|
Gain
|
|
Programmable from 1 to 5,000 in 1, 2, 3, 5 steps with ±0.05% accuracy.
|
|
Gain Stability
|
|
±0.01% for 30 days, ±0.004%/ºC.
|
|
Gain Linearity
|
|
±0.01% for gains <1,000, ±0.02% for gain 1,000 and higher.
|
|
Impedance
|
|
50 Megohms, shunted by 500 pF.
|
|
Input Protection
|
|
±50 Volts, differential or ±350 Volts common mode without damage.
|
|
Common Mode
|
|
80 dB plus gain in dB to 120 dB for balance input and 110 dB for a 350 Ohm source unbalanced, ±300 Volts, DC to 60 Hz.
|
|
CM Voltage
|
|
Common Mode ±300 Volts operating, ±350 Volts without damage.
|
|
Zero (Amplifier)
|
|
Automatic zero ±1.0 mV. Stability is ±5 µV RTI, ±1 mV RTO, ±1 µV RTI/ºC, ±0.2 mV RTO/ºC. Short term ±2 µV RTI, ±0.4 mV RTO.
|
|
Zero Stability
|
|
±1 µV/°C RTI, ±0.2 mV/°C RTO or (±1 µV RTI, ±0.2 mV RTO) /°C
|
|
Source Current
|
|
±25 nA, ±0.05 nA/ºC.
|
|
Noise (10 kHz)
|
|
2.0 µV RTI plus 0.3 mV RTO, RMS.
|
|
Bandwidth
|
|
6160: 50 kHz for gains <1,000, 20 kHz for gains greater than 1,000. 6160HF: 100 kHz for gains <1,000, 50 kHz for gains greater than 1,000.
|
|
Slew Rate
|
|
5 V/µS
|
|
|
|
|
|
|
|
|
Voltage Substitution
|
|
Voltage substitution, signal from external calibration source is applied to the amplifier input. Programmable attenuator with steps of 1, 0.1 and 0.01, ±0.02% accuracy. Output of the attenuator is provided for calibration.
|
|
Shunt Calibration
|
|
Shunt Calibration based on capability of Installed Features Card. FC1: Two steps, single shunt, internal or external. FC2: Programmable resistive "DAC" shunt, 16-bit resolution. FC5: Four-step, single shunt, external. FC11: Four-step, double shunt, external.
|
|
Zero Calibration
|
|
Amplifier input disconnected and shorted.
|
|
|
|
|
|
|
Ordering Information
|
|
6160-PF4-BE6 - 2-Channel Transducer Amp, 4-Freq 6-Pole Bessel.
6160-PF4/10K-BE4 - 2-Channel Transducer Amp, 4-Pole PF 4 Hz to 10 kHz 4-Pole Bessel.
6160HF-PF4-BE6 - 2-Channel Transducer Amp, 4-Freq 6-Pole Bessel.
6160HF-P10/20K-BE4 - 2-Channel Transducer Amp, 4-Pole PF 10 Hz to 20 kHz 4-Pole Bessel
6060-FC1 - 2-step local/remote shunt.
6060-FC2 - DAC shunt.
6060-FC5 - 4-step single shunt.
6060-FC7 - Fixed AC Coupling.
6060-FC8 - RTD input.
6060-FC11 - 4-step double shunt.
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|